beautypg.com

Hanna Instruments HI 2214 User Manual

Page 12

background image

12

FOUR, THREE OR TWO-POINT CALIBRATION

• Proceed as described in “FIVE-POINT CALIBRATION” section.

• Press

CAL after the fourth, third or second calibration point was

confirmed. The instrument will memorize the calibration data and
return to measurement mode.

ONE-POINT CALIBRATION

Two SETUP selectable options are available: “

Pnt” and “OFFS”.

If the “

Pnt” option is selected, the new calibration point overrides an

existing one. The adjacent slopes will be reevaluated.

If the “

OFFS” option is selected, an electrode offset correction is

performed. The adjacent slopes will remain unchanged.

• Proceed as described in “FOUR, THREE or TWO-POINT CALIBRATION”

section.

• Press

CAL after the first calibration point was confirmed. The

instrument will memorize the one-point calibration data and return
to measurement mode.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected

buffer, the “

READY” tag will be displayed and the

CFM” tag will blink.

• Press

CFM to confirm calibration.

• The calibrated value is then displayed on the

primary LCD and the fifth expected value on the
secondary LCD.

• After the fourth calibration point is confirmed, submerse the pH

electrode and the temperature probe approximately 3 cm (1¼”)
into the next buffer solution and stir gently. The temperature
probe should be close to the pH electrode.

• If necessary, press the

ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected

buffer, the “

READY” tag will be displayed and the

CFM” tag will blink.

• Press

CFM to confirm the fifth calibration point.

• The instrument stores the calibration value of the

fifth point and returns to normal measurement mode.

This manual is related to the following products: