Performing Acquisitions

60

IQS-1700

Selecting the Sampling Type

The acquisition rate applies to all channels when using a multichannel
power meter.

You can also change the continuous sampling rate in the Instrument
function tab in the same manner.

If you select Single, select the rate to use with the arrow buttons.

This rate is a divider of the maximum rate (5208 Hz). For example,
5208 Hz has a divider of 1, 2604 has a divider of 2 and so forth. To add
or remove dividers from the list of available values, see Editing the
Frequency Divider List
on page 61. The acquis
ition rate applies to all
channels when using a multichannel power meter.

If you select User Defined, select the rate by entering a value in the 1/
list, then select a time unit using the arrow buttons. Remember to click

to confirm your settings.

Note: Use a lower sampling rate with averaging set to active (see Averaging

Measurements on page 51) for greater repeatability when measuring very
low power.