beautypg.com

Sense[1..n]:average:count – EXFO IQS-1700 High Performance Power Meter for IQS-500/600 User Manual

Page 146

background image

SCPI Command Reference

138

IQS-1700

Product-Specific Commands—Description

:SENSe[1..n]:AVERage:COUNt?

Description

This query returns the number of measurements

used to perform an averaged measurement on

the specified channel.

Syntax

:SENSe[1..n]:AVERage:COUNt?[MINimu

m|MAXimum|DEFault]

Parameter(s)

Parameter 1:
The program data syntax for the first parameter is

defined as a

element. The allowed

DATA> elements for this parameter are:

MINimum|MAXimum|DEFault.

MINimum is used to retrieve the instrument's

smallest supported value.

MAXimum is used to retrieve the instrument's

greatest supported value.

DEFault is used to retrieve the instrument's

default value.

Response Syntax