A technical specifications, Mportant, High performance power meter – EXFO IQS-1700 High Performance Power Meter for IQS-500/600 User Manual
Page 113: Specifications, Iqs/ohs-1700 series), General specifications

High Performance Power Meter
105
A Technical Specifications
I
MPORTANT
The following technical specifications can change without notice.
The information presented in this section is provided as a reference
only. To obtain this product’s most recent technical specifications,
visit the EXFO Web site at www.exfo.com.
SPECIFICATIONS
a
(IQS/OHS-1700 SERIES)
Optical heads must be operated with the IQS-1710/1720/1740 High-Performance Interface Module.
Model
IQS-1712X/1722X/1742X
IQS-1713/1723/1743
OHS-1713-UH
Number of detectors
1/2/4
1/2/4
1
Detector type
GeX
InGaAs
InGaAs and integrating cavity
Detector size
2 mm detector
1 mm detector
9 mm input aperture
Wavelength range (nm)
800 to 1660
800 to 1700
930 to 1660
Power range (dBm)
b, h
25 to
<50
8 to
<87
37 to
<55
Uncertainty
±(5 % + 10 nW)
e, h
±(5 % + 2 pW)
f, h
±(4 % + 3 nW)
g, h
Polarization-dependent responsivity (dB)
c
N/A
N/A
(0 dBm to
<50 dBm)
±0.008 typ.
Linearity
d
±0.015 dB (10 dBm to
<30 dBm)
±0.015 dB (5 dBm to
<55 dBm)
±0.11 dB (35 dBm to 30 dBm)
±0.05 dB (30 dBm to 5 dBm)
±0.015 dB (5 dBm to −22 dBm)
Power resolution (dB)
h
0.001 (25 dBm to
<20 dBm)
0.001 (8 dBm to
<50 dBm)
0.001 (37 dBm to
<25 dBm)
Wavelength resolution (nm)
0.01
0.01
0.01
Stabilization time (ms)
0.3
0.3
0.3
Sampling rate (sample/s/channel)
up to 5208
up to 5208
up to 5208
Fiber type (μm)
5/125 to 62.5/125
NA
) 0.3
5/125 to 62.5/125
NA
) 0.3
5/125 to 62.5/125
NA
) 0.4 (with BFA-3000)
Notes
a. At 1550 nm (unless otherwise specified), with an FC angled connector and a warmup time
of 20 minutes, followed by an offset nulling.
b. From 18 °C to 28 °C.
c.
At 23 °C ± 3 °C, constant wavelength (1550 nm), constant power and with an FC non-angled
connector.
d. At constant temperature in the 0 °C to 40 °C range; nulling required.
e.
At 23 °C ± 1 °C with an FOA-322 and an FC non-angled connector, between 1000 nm and
1650 nm. Add 1 % to uncertainty below 1000 nm, and 3 % over 1650 nm.
f.
At 23 °C ± 1 °C with an FOA-322 and an FC non-angled connector, between 1000 nm and
1640 nm. Add 1 % to uncertainty below 1000 nm, and 6 % over 1640 nm.
g.
At 23 °C ± 1 °C with an FOA-322 and an FC angled connector, between 1290 nm and 1340 nm,
and between 1420 nm and 1640 nm. Add 2 % to uncertainty below 1000 nm, 1 % between
1370 nm and 1420 nm, and 5 % over 1640 nm. All uncertainties valid on the day of calibration.
Wavelength must not be equal to any water absorption line.
h. Averaging time of 1 s.
i. For optical power of > 35 dBm, maximum operating temperature is 30 °C. With the FOA-396,
maximum operating temperature is 25 °C.
j. From 0 °C to 40 °C.
GENERAL SPECIFICATIONS
IQS-1710/1720/1740
IQS-1713/1723/1743
IQS-1712X/1722X/1742X
OHS-1713-UH
Size (H x W x D)
125 mm x 36 mm x 282 mm
(4
15
/
16
in x 1
7
/
16
in x 11
1
/
8
in)
42 mm x 79 mm x 190 mm
(1
5
/
8
in x 3
1
/
8
in x 7
1
/
2
in)
Weight
0.64 kg (1.4 lb)
0.5 kg (1.1 lb)
Temperature
Operating
i
Storage
0 °C to 40 °C
(32 °F to 104 °F)
<40 °C to 70 °C
(
<40 °F to 158 °F)
0 °C to 40 °C
(32 °F to 104 °F)
<40 °C to 70 °C
(
<40 °F to 158 °F)
Number of ports
1/2/4
1
Relative humidity
j
0 % to 80 %
non-condensing
0 % to 80 %
non-condensing
Remote control
With IQS-600: GPIB (IEEE-488.1, IEEE-488.2), Ethernet and RS-232.
Instrument drivers
LabVIEW™, SCPI commands and COM/DCOM library drivers.
Standard accessories
User guide, one fiber-optic adapter per channel, Certificate of Compliance and Certificate of Calibration.