SCPI Command Reference

170

IQS-1700

Product-Specific Commands—Description

:SENSe[1..n]:POWer:WAVelength?

Description

This query returns the currently selected

wavelength on the specified channel. If no

channel was specified, the default value used is

1.

Syntax

:SENSe[1..n]:POWer:WAVelength?[MINi

mum|MAXimum|DEFault]

Parameter(s)

Parameter 1:
The program data syntax for the first parameter is

defined as a

element. The allowed

DATA> elements for this parameter are:

MINimum|MAXimum|DEFault.

MINimum is used to retrieve the instrument's

smallest supported value.

MAXimum is used to retrieve the instrument's

greatest supported value.

DEFault is used to retrieve the instrument's

default value.

Response Syntax

Response(s)

Wavelength:
The response data syntax for is

defined as a

element.

This query returns the current wavelength in

meters for the specified channel.

Example(s)

SENS:POW:WAV?

See Also