beautypg.com

EXFO IQS-1700 High Performance Power Meter for IQS-500/600 User Manual

Page 137

background image

SCPI Command Reference

High Performance Power Meter

129

Product-Specific Commands—Description

:MMEMory[1..n]:ACQuisition:DURation:

MMAXimum:STATe

Description

This command is used to set the timer state of

the Min. Max. acquisition.

Syntax

:MMEMory[1..n]:ACQuisition:DURation:MMAXim

um:STATe

Parameter(s)

StateDurMinMax:
The program data syntax for

is defined as a

element. The special forms

ON and OFF are accepted on input for increased

readability. ON corresponds to 1 and OFF

corresponds to 0.

Changes the timer state for the Min. Max.

acquisition.

0 -Timer for the Min. Max. acquisition is disabled.

1 -Timer for the Min. Max. acquisition is enabled.

Example(s)

MMEM:ACQ:DUR:MMAX:STAT ON

See Also

MMEMory:ACQuisition:DURation:MMAXimum

MMEMory:ACQuisition:DURation:MMAXimum:ST

ATe?

MMEMory:ACQuisition:DURation

MMEMory:ACQuisition

INITiate:AUTO

INITiate:EXTRema