Daq/daqe/pxi-2005/2006/2016 ai data format, Table 4-3: bipolar analog input range and output, Digital code for daq/daqe/pxi-2205/2206 – ADLINK PXI-2208 User Manual
Page 48: Table 4-4: unipolar analog input range and output

36
Operation Theory
DAQ/DAQe/PXI-2005/2006/2016 AI Data Format
The data format of the acquired 16-bit A/D data is 2's Comple-
ment coding. Table 4-3 and Table 4-4 illustrate the valid input
ranges and the ideal transfer characteristics.
Description
Bipolar Analog Input Range
Digital
code
Full-scale Range
±10V
±5V
±2.5V
±1.25V
—
Least significant bit
305.2 µV
152.6 µV
76.3 µV
38.15 µV
—
FSR-1LSB
9.999695V 4.999847V 2.499924V 1.249962V
7FFF
Midscale +1LSB
305.2 µV
152.6 µV
76.3 µV
38.15 µV
0001
Midscale
0V
0V
0V
0V
0000
Midscale -1LSB
-305.2 µV
-152.6 µV
-76.3 µV
-38.15 µV
FFFF
-FSR
-10V
-5V
-2.5V
-1.25V
8000
Table 4-3: Bipolar Analog Input Range and Output Digital Code for DAQ/DAQe/
PXI-2205/2206
Description
Unipolar Analog Input Range
Digital
code
Full-scale Range
0V to 10V
0 to +5V
0 to +2.5V 0 to +1.25V
—
Least significant bit
152.6 µV
76.3 µV
38.15 µV
19.07 µV
—
FSR-1LSB
9.999847V 4.999924V 2.499962V 1.249981V
7FFF
Midscale +1LSB
5.000153V 2.500076V 1.250038V 0.625019V
0001
Midscale
5V
2.5V
1.25V
0.625V
0000
Midscale -1LSB
4.999847V 2.499924V 1.249962V 0.624981V
FFFF
Table 4-4: Unipolar Analog Input Range and Output Digital Code for DAQ/DAQe/
PXI-2205/2206