Sense : telecom : measure : type Page 94
Page 94
Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

8.3 Device-Specific Commands
: SENSe : TELecom : FILTer?
Command type
Response
Function
Restriction
Example of use
Sequential command
Makes a query of the filter setting.
HP'+LP, HPl'+LP setting is disabled when the Bit Rate is set at 2.5 G or 5 G.
> : SENSe : TELecom : FILTer?
< HPISLP
: SENSe : TELecom : MEASure : TYPE
Command type
Sequential command
Parameter
MANual........Manual measurement
SINGle.......... Single measurement
REPeat.......... Repeat measurement
Function
Performs the measurement mode setting.
Example of use
Sets the measurement mode to the repeat measurement.
> : SENSe ; MEASure ; TYPE REPeat
: SENSe : TELecom : MEASure : TYPE?
Command type
Sequential command
Response
MAN............. Manual measurement
SING............. Single measurement
REP............... Repeat measurement
Function
Makes a query of the measurement mode.
Example of use
> : SENSe : MEASure : TYPE?
< REP