A.3 Device-Specific Errors

A.3 Device-Specific Errors

The error codes, [-399 and -300J, indicate that the device encountered errors other than the command error, query error

and run time error. Device-specific errors include errors such as hardware and firmware errors and self-diagnosis error.

When a device-specific error takes place, bit 3 of the event status register is set.

Table A-3 Error Messages of Device-Specific Errors

Code

Message

Error Detection Condition

-310

System error

A system error has taken place.

-314

Save/recall memory lost

The local memory has been lost.

-315

Configuration memory lost

The resume memory has been lost.

N O T E :

The self-test includes a self test conducted at the power activation.