Section 5 Common IEEE488.2 Commands
^CLS
Clear Status Command
Parameter
Function
Example of use
None
Clears all the status composition excluding the output queue and MAV summary message
Flowever, the enable register and Transition filter are not cleared.
Clears both the output queue and MAV bit when *CLS is sent immediately after the GRAM MESSAGE TERMINATOR>, and at the same time, before the UNIT> element. > *CLS ^ E S E Standard Event Status Enable Command Parameter Function Example of use Integer value between 0 and 255. Set as the parameter the sum total of the bits desired to be enabled among the standard event status enable register. Set 1 for enable and 0 for disable. See the Section for the "Status Byte" for the composition of the register of this device. Sets and clears the standard event status enable register. Sets bits 2 and 4 of the standard event status enable register. > *ESE20 ^ESR? Standard Event Status Register Query Response Function Example of use Integer value between 0 and 255. Set as the response the sum total of the bits of the standard event status register. See the Section for the "Status Byte" for the composition of the register of this device. Inquires about the current value of the standard event status register. When a command error is found. > *ESR? < 32 5-4 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888)