Section 8 Device Message Details
: SENSe : TELecom : RANGe
Command type
Parameter
Function
Example of use
Sequential command
UI4 ...............
4 UI
pp
UIl ................ 1 UI
pp
Sets the measurement range of receiving jitter.
Sets the measurement range of receiving jitter at 1
UI
pp
.
> : SENSe : TELecom : RANGe UIl
: SENSe : TELecom : RANGe?
Command type
Response
Function
Example of use
Sequential command
Makes a query of receiving jitter measurement.
> : SENSe : TELecom : RANGe?
< UIl
: SENSe : TELecom : FILTer
Command type
Sequential command
Parameter
LP.................. LP
(80 M)
HPILP............HPl+LP
(10 K to 80 M)
HPISLP......... HPl’+LP
(20 K to SOM)
HP2LP...........HP2+LP (4 M to 80 M)
HP3LP...........HP3+LP (6 M to 80 M)
HPLP.............HP+LP
(12 K to SOM)
HPSLP...........HP'+LP
(50 K to 80 M)
Function
Performs filter setting.
Restriction
HP’+LP, HPl'+LP setting
is disabled
when the Bit Rate is set at 2.5 G or 5 G.
HP3+LP can be set only for Option 07.
Example of use
Sets the filter at HP'+LP, HPl’+LP.
> ; SENSe : TELecom : FILTer HPSLP
8-16