9 indexes, 1 explanation of the messages, 1 error messages – YSI IQ SensorNet FDO Sensor User Manual
Page 43: Indexes -1, Explanation of the messages -1 9.1.1, Error messages -1, 9indexes

FDO
®
70x IQ ...
Indexes
ba76014e03
05/2014
9 - 1
9
Indexes
9.1
Explanation of the messages
This chapter contains a list of all the message codes and related
message texts that can occur in the log book of the IQ
S
ENSOR
N
ET
system for the FDO
®
70x IQ ... sensor.
9.1.1
Error messages
Information on the contents and structure of the log book,
and how to call it up, is given in the
L
OG
BOOK
chapter of the
IQ
S
ENSOR
N
ET
system operating manual.
The last three digits of the message code identify the
source of the message:
334 = SC FDO 700 (sensor cap SC-FDO
®
700)
335 = SC FDO 701 (sensor cap SC-FDO
®
701)
531 = FDO700IQ (sensor basis / component class,
adapter ADA)
Message code
Message text
EA133x
Meas. range exceeded or undercut
* Check process
* Select other meas. range
EA233x
Sensor temperature too high!
* Check process and application
EA333x
Sensor temperature too low!
* Check process and application
EAP33x
Measurement interfered
* SensCheck: Sensor cap is missing, leaky, depleted, or defective
* Clean sensor and space between cap and sensor according to
op.instructions
* Screw on sensor cap securely
* Replace sensor cap
EAP531
Measurement interfered
* SensCheck: Sensor cap is missing, leaky, depleted, or defective
* Clean sensor and space between cap and sensor according to
op.instructions
* Screw on sensor cap securely
* Replace sensor cap