YSI IQ SensorNet FDO Sensor User Manual
Page 23
FDO
®
70x IQ ...
Measuring / Operation
ba76014e03
05/2014
4 - 5
Canceling the user
calibration
As long as the determination of the calibration data has not yet been
started (step 6), you can quit the calibration routine with
The running determination of calibration data (after pressing
step 6) can be aborted as follows:
After the user calibration was canceled the sensor again works with the
calibration data that was used before the canceled user calibration.
Possible results of the
user calibration
The calibration data is evaluated by the system. A calibration can have
the following results:
1
Open the setting table (see section 3.4).
2
In the Calibration menu item, select the abort setting and then
quit the setting table with Save and quit.
Display
Explanation
Measured value display
Sensor was successfully calibrated.
The calibration data can be viewed in the
calibration history (section 4.2.4).
"----"
The sensor could not be calibrated.
The sensor is blocked for further measure-
ment.
Notes on possible causes are given in the
log book of the sensor.