YSI IQ SensorNet FDO Sensor User Manual
Page 21

FDO
®
70x IQ ...
Measuring / Operation
ba76014e03
05/2014
4 - 3
Canceling the check
As long as the determination of the relative slope has not yet been
started (step 6), you can quit the check with
The running determination of the relative slope (after pressing
step 6) can be aborted as follows:
Evaluation
The result of the check is entered in the log book of the sensor. If the
relative slope is outside the tolerance range (0.90 ... 1.10), the check is
assessed as erroneous.
7
Put the sensor in the measuring position again.
8
Wait for the measured value to be largely stable (temperature
adjustment).
9
Switch off the maintenance condition.
1
Open the setting table (see section 3.4).
2
In the Test menu item, select the abort setting and then quit the
setting table with Save and quit.
The information recorded in the log book entry is partly
identical with that of the calibration history. The calibration
history is described in detail in section 4.2.4 C
- 3100 (50 pages)
- 3200 (86 pages)
- 5000 (70 pages)
- 5200 (148 pages)
- 5200A (128 pages)
- 5200A (121 pages)
- 5400 (127 pages)
- 5239 Dissolved Oxygen Probe (23 pages)
- 55 (26 pages)
- 550A (18 pages)
- 556 MPS - Quick Start Guide (4 pages)
- 58 (40 pages)
- 5905 (22 pages)
- 6600V2 - User Manual (379 pages)
- 60 (33 pages)
- 63 (46 pages)
- 650 MDS (72 pages)
- 9300 Photometer (11 pages)
- 9500 Photometer (143 pages)
- 9500 Photometer (18 pages)
- CR 3200 Thermoreactor (42 pages)
- CR 4200 Thermoreactor (42 pages)
- CR2200 Thermoreactor (36 pages)
- DO200 (40 pages)
- EC300 (40 pages)
- EC300A (40 pages)
- EC30A (12 pages)
- IFL 700 IQ (46 pages)
- IQ SensorNet 2020 XT Terminal (196 pages)
- IQ SensorNet MIQ/A-MOD Module (42 pages)
- IQ SensorNet MIQ/A-PR Module (36 pages)
- IQ SensorNet MIQ/IC2 Module (30 pages)
- IQ SensorNet MIQ/IF232 Module (18 pages)
- IQ SensorNet MIQ/JB Module (18 pages)
- IQ SensorNet MIQ/JBR Module (20 pages)
- IQ SensorNet MIQ/PS Module (22 pages)
- IQ SensorNet MIQ/R6 Module (56 pages)
- IQ SensorNet NitraLyt Sensor (64 pages)
- IQ SensorNet SensoLyt Sensor 700 IQ (SW) (46 pages)
- IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
- IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
- IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
- MultiLab 4010-1 (108 pages)
- MultiLab 4010-2 (138 pages)