Commissioning
ba76014e03
05/2014
3 - 1
3
Commissioning
3.1
Scope of delivery
Sensor, consisting of sensor basis and sensor cap
Operating manual.
3.2
Installation
Safety guidelines for
installation in a
hazardous location
(H-models only)
Connection cable
A sensor connection cable of the SACIQ or SACIQ SW type is required
to connect the sensor. The cable is available in different lengths. Com-
pared to the standard model SACIQ, the SACIQ SW sensor connection
cable is optimized regarding its corrosion resistance in seawater and
brackish water and adapted for use in conjunction with the FDO 70x
IQ SW. Information on this and other IQ
S
ENSOR
N
ET
accessories is
given in the YSI catalog and on the Internet.
DANGER
Explosion hazard.
Only hazardous location rated sensor models
("Hazloc Sensor") must be used in hazardous
locations. Verify the rating on the product name
plate on the sensor shaft.
The associated apparatus MIQ/BB1 (YSI part no.
207001Y) has to be used for connection. Refer to the
following documents for proper installation:
Control drawing, YSI document no. 28123
(see MIQ/BB1 operating manual)
MIQ/BB1 operating manual
All applicable electrical code regulations.
DANGER
Explosion hazard.
For hazardous location use only the cable types list-
ed in the control drawing must be used.
How to connect the SACIQ (SW) sensor connection cable
to the terminal strip of a MIQ module is described in chapter
3
I
NSTALLATION
of the IQ
S
ENSOR
N
ET
system operating
manual.