6 what to do if, What to do if -1, 6what to do if – YSI IQ SensorNet FDO Sensor User Manual
Page 33

FDO
®
70x IQ ...
What to do if...
ba76014e03
05/2014
6 - 1
6
What to do if...
Sensor does not appear
in the measured value
display and list of
sensors
Measured value
implausible
Measured value
fluctuating heavily
Cause
Remedy
– Sensor cap not mounted or
defective
– Sensor cap (see section
– Replace a defective sensor
cap as necessary
Cause
Remedy
– Coating on sensor cap
– Clean the outside of the sen-
– Service life of the sensor cap
over
– Check the sensor (see sec-
– Replace the sensor cap as
necessary (see section
– Dirt inside the sensor cap and
sensor head.
– Fixing ring not properly tight-
ened.
– Sensor cap untight or defec-
tive.
– Dismantle sensor cap
– Clean the insides of the sen-
sor cap and sensor head (see
section
– Mount the sensor cap cor-
rectly and tighten the fixing
ring as far as it will go (see
section
– Replace a defective sensor
cap as necessary
Cause
Remedy
– Coating on sensor cap
– Clean the outside of the sen-
– Service life of the sensor cap
over
– See point "Measured value
– Dirt inside the sensor cap and
sensor head.
– Fixing ring not properly tight-
ened.
– Sensor cap untight or defec-
tive.
– See point "Measured value