beautypg.com

5 reactivating previous calibration data, Reactivating previous calibration data – YSI IQ SensorNet FDO Sensor User Manual

Page 25

background image

FDO

®

70x IQ ...

Measuring / Operation

ba76014e03

05/2014

4 - 7

4.2.5

Reactivating previous calibration data

The FDO

®

70x IQ ... enables you to reactivate the last valid user cali-

bration or the factory calibration. Thus you can immediately go on
measuring if a calibration procedure failed or you suspect that the cali-
bration conditions were not optimally met.

Reactivating calibration

data

Reactivating old calibration data is only a temporary solu-
tion. Please take into account that the sensor might thus
possibly provide incorrect measured values. Please make
sure the sensor works correctly by carrying out a function
check and/or recalibration.

1

Open the setting table (see section 3.4).

2

In the Calibration menu item, select the User calibration or Fac-
tory calibration
setting and then quit the setting table with Save
and quit
.