beautypg.com

4 fdo® 70x iq ... setting table – YSI IQ SensorNet FDO Sensor User Manual

Page 16

background image

3 - 4

ba76014e03

05/2014

Commissioning

FDO

®

70x IQ ...

3.4

FDO

®

70x IQ ... setting table

Setting

Selection/values

Explanation

Measuring mode

Concentration
Saturation

Unit of the measured value on the measured
value display.

Measuring range
Concentration

0 ... 20.00 mg/l
0 ... 20.00 ppm

These measuring ranges are available for
selection.

Measuring range Sat-
uration

0 ... 200.0 %

The measuring range is set permanently.

Response time t90

 150 .. 300 s

(with SC-FDO

®

700)

or
 60 ... 300 s

(with SC-FDO

®

701)

Response time of the signal filter.
Depending on the sample matrix, the mea-
sured values may fluctuate more or less.
A signal filter in the sensor reduces the fluc-
tuation range of the measured value. The
signal filter is characterized by the Response
time t90
. This is the time after
which 90 % of a signal change is displayed.

The setting range depends on the sensor
cap type.

Calibration

Factory calibra-

tion

User calibration

active
abort

Determines which calibration data the mea-
sured value calculation will be based on. The
active calibration is displayed in the calibra-
tion history.

The selection, User calibration is only dis-
played if valid data of a User calibration is
stored in the sensor.

active indicates that the sensor is being cali-
brated.

If abort is selected, the active calibration is
canceled the next time the setting table is
quit with Save and quit.

Test

active
abort

active indicates that the sensor is being
checked.

If abort is selected, the active check is can-
celed the next time the setting table is quit
with Save and quit.

Temperature mode

°C
°F

Unit of the measured temperature value
(Celsius, Fahrenheit).