4 fdo® 70x iq ... setting table – YSI IQ SensorNet FDO Sensor User Manual
Page 16
3 - 4
ba76014e03
05/2014
Commissioning
FDO
®
70x IQ ...
3.4
FDO
®
70x IQ ... setting table
Setting
Selection/values
Explanation
Measuring mode
Concentration
Saturation
Unit of the measured value on the measured
value display.
Measuring range
Concentration
0 ... 20.00 mg/l
0 ... 20.00 ppm
These measuring ranges are available for
selection.
Measuring range Sat-
uration
0 ... 200.0 %
The measuring range is set permanently.
Response time t90
150 .. 300 s
(with SC-FDO
®
700)
or
60 ... 300 s
(with SC-FDO
®
701)
Response time of the signal filter.
Depending on the sample matrix, the mea-
sured values may fluctuate more or less.
A signal filter in the sensor reduces the fluc-
tuation range of the measured value. The
signal filter is characterized by the Response
time t90. This is the time after
which 90 % of a signal change is displayed.
The setting range depends on the sensor
cap type.
Calibration
Factory calibra-
tion
User calibration
active
abort
Determines which calibration data the mea-
sured value calculation will be based on. The
active calibration is displayed in the calibra-
tion history.
The selection, User calibration is only dis-
played if valid data of a User calibration is
stored in the sensor.
active indicates that the sensor is being cali-
brated.
If abort is selected, the active calibration is
canceled the next time the setting table is
quit with Save and quit.
Test
active
abort
active indicates that the sensor is being
checked.
If abort is selected, the active check is can-
celed the next time the setting table is quit
with Save and quit.
Temperature mode
°C
°F
Unit of the measured temperature value
(Celsius, Fahrenheit).