4 instrument safety, 1 general instrument safety, 2 hazardous location ratings (h models) – YSI IQ SensorNet FDO Sensor User Manual
Page 37: 5 electrical data, Instrument safety -3 7.4.1, General instrument safety -3, Hazardous location ratings (h models) -3, Electrical data -3

FDO
®
70x IQ ...
Technical data
ba76014e03
05/2014
7 - 3
7.4
Instrument safety
7.4.1
General instrument safety
7.4.2
Hazardous location ratings (H models)
Applicable directives
and standards
In addition to the standards listed in chapter 7.4.1, the FDO
®
70x IQ H
conforms to the following directives and standards:
ANSI/ISA-12.12.01
CAN/CSA C22.2 # 213
The FDO
®
70x IQ H is a non-incendive electrical equipment for use in
Class I Division 2 hazardous locations.
Hazardous location
rating
7.5
Electrical data
Applicable norms
– UL 61010-1
– CAN/CSA C22.2 # 61010.1
FDO
®
700 IQ H
FDO
®
701 IQ H
Class I Division 2
Groups A, B, C, D T6
Class I Zone 2
Group IIC T6
Associated apparatus
MIQ/BB1, YSI part no. 207001Y
Control Drawing
YSI document no. 28123
Rated voltage
(non-H models)
24 V DC
via the IQ
S
ENSOR
N
ET
(for more
details, see chapter
T
ECHNICAL
DATA
of the IQ
S
ENSOR
N
ET
sys-
tem operating manual)
Voltage range
(H models)
14 V DC ... 24 V DC via the MIQ/
BB1 barrier box (=associated
apparatus)
Power consumption
0.7 W
Protective class
III