Technical data

ba76014e03

05/2014

7 - 5

7.7

Technical data of the SC-FDO

®

701

Measuring ranges and

resolutions

D. O. partial pressure 0 ... 400 hPa

Measurement accuracy

Repeatability

± 0.05 mg/L (ppm) according to EN ISO 15839

Response time

t

90

(90 % of the final display value after) < 60 s according to

EN ISO 15839
t

95

(95 % of the final display value after) < 80 s according to

EN ISO 15839

Minimum approach flow

No incident flow required

Interferences

Non-sensitive to against hydrogen sulfide, chlorine, and ionogenic
substances

Allowed temperature

range

Allowed pH range of the

test sample

4 ... 12

Calibration procedure

Factory calibration by means of IQMC (Intelligent Membrane Calibra-
tion) procedure. User calibration possible in water vapor-saturated air
or air-saturated water.

Material

PMMA, PVC and silicone

Working life

6 months guaranteed for authorized use under the specified environ-
mental conditions

Measuring
mode

Adjustable measuring range

Resolution

D. O. concentra-
tion

0 ... 20.00 mg/L
0 ... 20.00 ppm

0.01 mg/L
0.01 ppm

D. O. saturation

0 ... 200.0 %

0.1 %

In the range

Accuracy

< 1 mg/L (ppm)

± 0.05 mg/L (ppm)

> 1 mg/L (ppm)

± 0.1 mg/L (ppm)

Measuring medium

-5 °C ... + 40 °C (23 ... 104 °F)

Storage/transport

- 10 °C ... + 40 °C (-33 ... 104 °F)