Measuring / Operation
FDO
®
70x IQ ...
4.2.4
Calibration history
Calibration history
(available in the
IQ
S
ENSOR
N
ET
systems
184 XT and 2020 XT
only)
Fig. 4-1
Calibration historyFDO
®
70x IQ ...
The calibration history contains the following information:
Date
Calibration date (Factory = factory calibration)
Rel.slope
Relative slope (non-dimensional)
Intens.
Intensity:
+
: sufficient
-
: too
low
Res
Evaluation of the calibration
o.k.
: Calibration
successful.
The new calibration data was taken over for measure-
ment.
Error
: Calibration
unsuccess-
ful.
Invalid calibration data were discarded. Measurement
was continued with the last valid values.
The calibration history is stored in the sensor basis. If
another sensor cap (with different series number) is
mounted, the calibration history is deleted and cannot be
restored afterwards.
Currently active
calibration
Chronological list of
the last user cali-
brations