Fault
Possible cause
Remedy
❑ Pump does not achieve
its ultimate vacuum or
usual pumping speed.
➨ Outgassing substances or
vapour generated in the proc-
ess?
✔ Check process parameters.
➨ Diaphragms or valves defec-
tive?
✔ Replace diaphragms and/or
valves.
➨ Pump too hot?
✔ Allow pump to cool down.
Determine and eliminate the
cause of overheating.
➨ Pressure below ”Minimum” in
Auto mode?
✔ Adapt switch off pressure
(”Minimum”) if necessary.
❑ Pump too noisy.
➨ Loud exhaust noise?
✔ Connect hose or silencer to
pump outlet.
➨ Diaphragm crack or dia-
phragm clamping disc loose?
✔ Perform maintenance.
➨ Other than above mentioned
causes?
✔ Contact local distributor.
❑ Pump seized.
✔ Contact local distributor.
A service manual with exploded view drawings, spare parts list and directions for
repair is available on request.
☞
The service manual is intended for trained service people only.