VACUUBRAND PC 3012 VARIO User Manual
Page 11
Documents are only to be used and distributed completely and unchanged. It is strictly the users’ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999164 / 19/05/2009
page 11 of 75
Type
PC 3002
VARIO
PC 3003
VARIO
PC 3004
VARIO
Maximum pumping speed (ISO
21360)*
m
3
/h
2.8
2.8
4.6
Ultimate vacuum (absolute) without
gas ballast**
mbar
7
0.6
1.5
Ultimate vacuum (absolute) with gas
ballast**
mbar
12
2
3
Maximum permissible inlet pressure
(absolute)
bar
1.1
Maximum permissible outlet pressure
(absolute)
bar
1.1
Maximum permissible pressure (abso-
lute) at gas ballast valve
bar
1.2
Permissible ambient temperature
storage / operation
°C
-10 to +60 / +10 to +40
Permissible relative atmospheric
moisture during operation (no conden-
sation)
%
30 to 85
Power (electric)
W
575
No-load speed
min
-1
30 - 2400
Maximum permissible range of supply
voltage (
±10% )
Attention: Observe specifications of
rating plate!
100-120 V~ 50-60 Hz
200-230 V~ 50-60 Hz
Maximum nominal current draw at:
100-120 V~ 50/60 Hz
200-230 V~ 50/60 Hz
A
A
3.2
1.4
6.3
2.5
Device fuse
slow blow fuse 4A (200-230V) / 8A (100-120V)
Motor protection
thermal cutout
Degree of protection IEC 529
IP 20
Inlet
hose nozzle NW 10 mm
Outlet
hose nozzle NW 10 mm
Coolant connections (waste vapour condenser)
hose nozzle NW 6-8 mm
Maximum permissible pressure of
coolant at waste vapour condenser
bar
6 (absolute)
Permissible range of coolant tempera-
ture
°C
-15 to +20
Dimensions L x W x H approx.
mm
422 x 258 x 516
Weight approx.
kg
18.4
21.0
* Pumping speed of diaphragm pump
** Ultimate vacuum at setting ”Pump down” at speed ”HI”