VACUUBRAND PC 3012 VARIO User Manual
Page 12
Documents are only to be used and distributed completely and unchanged. It is strictly the users’ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999164 / 19/05/2009
Type
MV 10C VARIO-B
PC 3010 VARIO
MD 12C VARIO-B
PC 3012 VARIO
Maximum pumping speed (ISO 21360)*
m
3
/h
8.6
10.0
Ultimate vacuum (absolute) without gas
ballast**
mbar
0.6
2
Ultimate vacuum (absolute) with gas bal-
last**
mbar
9
9
Maximum permissible inlet pressure
(absolute)
bar
1.1
Maximum permissible outlet pressure
(absolute)
bar
1.1
Maximum permissible pressure (absolute)
at gas ballast valve
bar
1.2
Permissible ambient temperature
storage / operation
°C
-10 to +60 / +10 to +40
Permissible relative atmospheric moisture
during operation (no condensation)
%
30 to 85
Power (electric)
W
575
No-load speed
min
-1
30 - 2400
Maximum permissible range of supply
voltage (
±10% )
Attention: Observe specifications of
rating plate!
100-120 V~ 50-60 Hz
200-230 V~ 50-60 Hz
Maximum nominal current draw at:
100-120 V~ 50/60 Hz
200-230 V~ 50/60 Hz
A
A
8
3.5
Device fuse
slow blow fuse 4A (200-230V) / 8A (100-120V)
Motor protection
thermal cutout
Degree of protection IEC 529
IP 20
Inlet Pump
Pumping unit
small flange NW 25
small flange NW 25 or hose nozzle NW 15
Outlet
hose nozzle NW 10 mm
Coolant connections (waste vapour condenser)
hose nozzle NW 6-8 mm
Maximum permissible pressure of
coolant at waste vapour condenser
bar
6 (absolute)
Permissible range of coolant temperature
°C
-15 to +20
Dimensions L x W x H approx.
Pump
Pumping unit
mm
505 x 237 x 303
645 x 365 x 600
Weight approx.
Pump
Pumping unit
kg
27.0
35.8
* Pumping speed of diaphragm pump
** Ultimate vacuum at setting ”Pump down” at speed ”HI”
We reserve the right for technical modifications without prior notice!