Continuous pumping with pressure and time settings

• Operation on demand of a speed controlled NT VARIO / VARIO-B pump

Preselections

Use the selection knob to select the parameters. All parameters can be altered even while operation

control is running.

Speed: Preselection of the motor speed for pump down. The selection ”HI” effects the maximum

speed and best ultimate vacuum of the pump (with automatic speed reduction at ultimate vacuum).

The ”Speed” is adjustable in a range of 1-100% and to ”HI”.

Minimum: The controller switches the pump off once the preset value for ”Minimum” has been

reached.

Minimum” is adjustable in a range of 1-1060 mbar or can be set to ”Off”.

A preset ”Duration” (process time) is without effect, if the process is stopped due to a preset ”Mini-

mum” before ”Duration” is reached.

Delay: Delay” determines the time the coolant valve is running on after the process has been

stopped.

The ”Delay” is adjustable in a range of 1-300 minutes or can be set to ”Off”.

Duration: Duration” determines the total process time since control start.

The process time is adjustable between 1-1440 minutes (24 h) or can be set to ”Off”. ”Off” indicates

that no endpoint for pump down is determined.

If a ”Duration” is preset, the controller switches off the pump when the preset process time is reached,

also if a preset ”Minimum” is still not reached.

If neither ”Minimum” nor ”Duration” are preset, process control has to be stopped by pressing key

”STOP”.

The screen-shot shows the factory-set values.

Pump down

Speed

HI

Minimum Off

Delay

Off

Duration Off

- - - - - - Graphic - - - - - - -

- - - - - - - Back- - - - - - - -

00:00:00

When selecting ”Graphic” the display shows a pressure vs. time curve.

The timeline in the diagram adapts automatically to the process time.

Press the selection knob twice to return to the standard display.

10 mbar

00:03:50

Function Pump down

Pump down