The controller CVC 3000 can be adapted to the specific application by choosing the appropriate function

depending on the connected components and the requirements of the application.

Automatic detection of the components

When switching on the controller the actual configuration of the connected components is checked

automatically.

Connected components (pump NT VARIO / VARIO-B, VSK 3000, valves) are detected automatically

and used and supervised until the controller is switched off. Switch off/on the controller to renew the

configuration.

The last mode of operation and the preselected values (e. g. for pressure, speed or time for switching

off) are stored.

In case of similar operation conditions it is possible to start immediately, if the preselections are chosen

appropriately.

The controller features

five functions and one configuration menu, see section ”Menu guide”. Each

of these functions includes specific preselection possibilities and is adapted automatically with regard to

the connected components.

Depending on the connected components (e.g. valves) some menu items might not be active!

Changing the function:

Switch controller on.

Press START/STOP to terminate control in case control is running.

Press key MODE.

Select function with knob and press to confirm.

Depending on the selected function the controller controls different components:

’’Pump down’’

• NT VARIO/VARIO-B pump depending on pressure and time preselections with continuous speed control

• Coolant valve

”Vac control”

• NT VARIO / VARIO-B pump with pinpoint precision according to pressure preselection

• Coolant valve

”Auto mode”

• NT VARIO / VARIO-B pump with pinpoint precision with fully automatic boiling point determination

and adaptation

• Coolant valve

’’Program’’

• Pump depending on pressure and time preselections or ”Auto mode”

• Coolant valve

• Venting valve

”VACUULAN”

• NT VARIO/VARIO-B pump depending on pressure and time preselections with continuous speed control

• Coolant valve

’’Configuration’’

Preselections for

• Adjustment of the pressure transducer

• Interface RS-232

• Sensors (configuration and switching between several sensors)

• Display (brightness and contrast of the display, language, sound)

• Autostart (automatic restart after power failure)

• Defaults (Resetting the controller is to factory set values)

The menu ”Configuration” is also accessible by pressing the selection knob while the start display is