Operation
Command Parameter Description
function
OUT_MODE X
1: continuous pumping
2: vacuum control without automatic
3: vacuum control with automatic
set vacuum
OUT_SP_1
XXXX
unit according to preselection
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)
set vacuum
OUT_SP_V
XXXX
unit according to preselection
with venting*
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)
selected frequency
OUT_SP_2
XX.X
motor frequency in Hz (01.0 to 60.0 in steps
of 0.5 Hz or 99.9 for “HI”)
pressure for restart
OUT_SP_3
XXXX
unit according to preselection
(VACUU•LAN)
(see OUT_SP_1)
delay
OUT_SP_4
XX:XX
hh:mm (hours:minutes)
vacuum for automatic OUT_SP_5 XXXX
unit according to preselection
switching off
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)
time for automatic
OUT_SP_6
XX:XX
hh:mm (hours:minutes)
switching off (VACUU•LAN)
starting process control START
stopping process control STOP
X
1: termination of process control
2: termination of process control and storage of
the current pressure as new set point
Remote operation**
REMOTE
X
0: set controller to local operation
1:
set
controller
to
remote
operation
* Pressure setting with venting is only possible in function ”Vacuum control” if an venting valve is
connected and configurated and vacuum control is started. The venting valve opens automatically
if the actual pressure is at least 10 mbar below the preset pressure. Automatic venting becomes
inactive if vacuum control is stopped (STOP or VENT), setting a pressure value using the command
OUT_SP_1 or if the function is changed. Activate the command OUT_SP_V again if necessary.
** If remote operation is selected or deselected, the user has to ensure that no dangerous status of
the system due to the change of the mode of operation can occur and to provide appropriate safety
measures, especially if selecting remote operation interferes with a locally operated active process.
Write commands ”CVC 2000”
2: RxD
5: Mass
9: +5V (Bluetooth)
3:
TxD
7:
RTS
4: DTR
8: CTS
Connection plug arrangement
driving venting valve
OUT_VENT
X
0: close venting valve (not automatically)
1: open venting valve
(process control stopped)
30: sensitivity: low
31: sensitivity: normal
32: sensitivity: high
(optional)