beautypg.com

Specifications – Oakworks PT Table User Manual

Page 34

background image

31

SPeCIfICATIOnS

Guidance and manufacturer’s declaration - electromagnetic immunity

The table is intended for use in the electromagnetic environment specified below. The customer or the

user of the table should assure that it is used in such an environment.

Immunity Test

IeC 60601 test level

Compliance level

electromagnetic

environment - guidance

electrostatic discharge

(eSD)

IeC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood,

concrete or ceramic tile.

If floors are covered with

synthetic material, the rela-

tive humidity should be at

least 30 %.

electrical fast transient/

burst

IeC 61000-4-4

±2 kV for power supply lines

±1 kV for input/output lines

±2 kV for power supply lines

±1 kV for input/output lines

Mains power quality should

be that of a typical com-

mercial or hospital environ-

ment.

Surge

IeC 61000-4-5

±1 kV differential mode

±2 kV common mode

±1 kV differential mode

±2 kV common mode

Mains power quality should

be that of a typical com-

mercial or hospital environ-

ment.

Voltage dips, short inter-

ruptions and voltage

variations on power supply

input lines IeC 61000-4-11

<5 % UT

(>95 % dip in UT)

for 0,5 cycle

40 % UT

(60 % dip in UT)

for 5 cycles

70 % UT

(30 % dip in UT)

for 25 cycles
<5 % UT

(>95 % dip in UT)

for 5 sec

<5 % UT

(>95 % dip in UT)

for 0,5 cycle

40 % UT

(60 % dip in UT)

for 5 cycles

70 % UT

(30 % dip in UT)

for 25 cycles

<5 % UT

(>95 % dip in UT)

for 5 sec

Mains power quality should

be that of a typical com-

mercial or hospital envi-

ronment. If the user of the

table requires continued

operation during power

mains interruptions, it is

recommended that the

table be powered from an

uninterruptible power sup-

ply or a battery.

Power frequency (50/60

Hz) magnetic field IEC

61000-4-8

3 A / m

3 A / m

Power frequency magnetic

fields should be at levels

characteristic of a typical

location in a typical com-

mercial or hospital environ-

ment.

NOTe U

T

is the a.c. mains voltage prior to application of the test level.

This manual is related to the following products: