Report Number: 100747045ATL-007

Issued:

Intertek

Identification, marking and documents (IEC 60601-1-2 (Ed. 3) Clause 5)

16.0

Guidance and manufacturer’s declaration –electromagnetic immunity


The PROLUX TABLE is intended for use in the electromagnetic environment specified below. The customer or the user of
the PROLUX TABLE should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic

environment

guidance

Electrostatic discharge
(ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air


Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.


Electrical fast
transient/burst

IEC 61000-4-4

±2 kV for power supply lines

±1 kV for input/output lines

±2 kV for power supply lines

±1 kV for input/output lines

Mains power quality should
be that of a typical
commercial or hospital
environment.


Surge

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

±1 kV differential mode

±2 kV common mode

Mains power quality should
be that of a typical
commercial or hospital
environment.


Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11


<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 sec


<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 sec


Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the PROLUX TABLE
requires continued
operation during power
mains interruptions, it is
recommended that the
PROLUX TABLE be
powered from an
uninterruptible power supply
or a battery.

Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8

N/A N/A


Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.

NOTE U

T

is the a.c. mains voltage prior to application of the test level.

Table 2

Data:

Page 70 of 70

EMC Report for Oakworks Inc. on the ProLux Table