background image

07/31/2012

Report Number: 100747045ATL-007

Issued:

Intertek

Identification, marking and documents (IEC 60601-1-2 (Ed. 3) Clause 5)

16.0

Guidance and manufacturer’s declaration –electromagnetic immunity

The PROLUX TABLE is intended for use in the electromagnetic environment specified below. The customer or the user of the
PROLUX TABLE should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance

level

Electromagnetic environment

guidance











Conducted RF
IEC 61000-4-6


Radiated RF
IEC 61000-4-3











3 Vrms
150 kHz to 80 MHz


3 V/m
80 MHz to 2,5 GHz










3 Vrms



3 V/m


Portable and mobile RF communications equipment should
be used no closer to any part of the [

ME

EQUIPMENT

or

ME

SYSTEM

], including cables, than the recommended

separation distance calculated from the equation applicable
to the frequency of the transmitter.

Recommended separation distance

P

d

2

.

1

=

P

d

2

.

1

=

80 MHz to 800 MHz

P

d

3

.

2

=

800 MHz to 2,5 GHz

where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).

Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey,

a

should be less than the

compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked
with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the PROLUX TABLE is used exceeds the applicable RF compliance level
above, the PROLUX TABLE should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the PROLUX TABLE.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.

Table 4

Data:

Page 69 of 70

EMC Report for Oakworks Inc. on the ProLux Table