beautypg.com

Intertek, Pd 2 . 1 p d 2 . 1, Pd 3 . 2 – Oakworks PerformaLift Series User Manual

Page 27: Data

background image

24

03/14/2013

Report Number: 101033247ATL-0

Issued:

Intertek

IEC 60601-1-2 (Ed.3) Clause 5 (IEC 60601-1-2 (Ed.3) Clause 5)

19.0

Guidance and manufacturer’s declaration –electromagnetic immunity

The PERFORMALIFT is intended for use in the electromagnetic environment specified below. The customer or the user of the

PERFORMALIFT should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance

level

Electromagnetic environment

guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2,5 GHz

3 Vrms

3 V/m

Portable and mobile RF communications equipment should

be used no closer to any part of the [

ME EQUIPMENT

or

ME

SYSTEM

], including cables, than the recommended

separation distance calculated from the equation applicable

to the frequency of the transmitter.

Recommended separation distance

P

d

2

.

1

P

d

2

.

1

80 MHz to 800 MHz

P

d

3

.

2

800 MHz to 2,5 GHz

where P is the maximum output power rating of the

transmitter in watts (W) according to the transmitter

manufacturer and d is the recommended separation

distance in metres (m).
Field strengths from fixed RF transmitters, as determined by

an electromagnetic site survey,

a

should be less than the

compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked

with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection

from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,

amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the

electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the

measured field strength in the location in which the PERFORMALIFT is used exceeds the applicable RF compliance level

above, the PERFORMALIFT should be observed to verify normal operation. If abnormal performance is observed, additional

measures may be necessary, such as re-orienting or relocating the PERFORMALIFT.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.

Table 4

Data:

Page 77 of 79

EMC Report for Oakworks Inc. on the Performalift

MAINTENANCE & SERVICE

MAINTENANCE

& SERVICE