beautypg.com

Canon CR-1 MARK II User Manual

Page 121

background image

8. Before calling the service technician (Error)

-113-

Continued on the following page >>

10005

1001

Failed to open a port.

Refer to 7.3.1 Retinal Camera Settings (see
page 77)
, and check the port number setting.

Disconnect the retinal camera communication
cable, then connect the cable again.

1003

Failed to set a port.

Refer to 7.3.1 Retinal Camera Settings (see
page 77)
, and check the port number setting.

1007

Power for the retinal camera is turned off or the
cable is improperly connected.

Check that the computer and retinal camera are
connected correctly.

1501

Error occurred during communications with the
retinal camera.
Failed to resend the command frame.

Turn the retinal camera’s power off and back on.

1502

Error occurred during communications with the
retinal camera.
Failed to resend the reply frame.

1503

Error occurred during communications with the
retinal camera.
Time-out for No Response

1504

Error occurred during communications with the
retinal camera.
Time-out for Incomplete Transmission

1603

Cameras are not attached to the main mount and
sub-mount of the retinal camera.

Check that the digital camera is connected correctly.

1701

Connected retinal camera model is different from
the model setting in this software.

Check the model of the retinal camera.

10007

4

JPEG File Saving Error

Check how much free memory is left on the drive
where the data is to be saved.

10010

1

Failed to log in.
Confirm User ID/Password.

Input the User ID and Password registered in the
user account (see page 100), and log in.

2

It cannot be set without administrator's privileges.

Log in using the User ID for which the Privilege has
been set to the Administrator (see page 100).

10011

2

This application is already running.

Use the software application which is running.

101

Set the screen resolution to 1024 × 768 or more.

Set the Windows screen resolution to 1024 × 768 or
more.

102

Set the screen color to 24 bits or more.

Set the Windows color quality to 24 bit or more.

110

Operation can be executed by a Windows user
with Administrator privileges.

Log into Windows using the name of the user
belonging to the Administrators group.

114

Essential data has not been set up.
Set up essential data.

Input the required items.

10012

27

Free space in the drive is not sufficient to capture
images.
Make up space in the drive by deleting studies in
Study Logs List.
[Drive:XX freesize: XXGByte(XXbyte)]
[Drive: XX]

Delete studies in the study log screen (see
page 38)
to increase the amount of available space
in the image storage drive.

29

Image cannot be captured due to insufficient space
to store captured image.
Make up space in the drive by deleting studies in
Study Logs List.
[Drive:XX freesize: XXGByte(XXbyte)]

30

Number of captured images exceeds the maximum
for the study.
No more image can be captured for the study.

End the study, and start a new study.

10019

7

MPPS Communication Error

Check that the MPPS server is operating properly.
Check the MPPS tab (see page 91) settings.

11

Study cannot be executed since a study having the
same Study Instance UID is running.

Check the study information which was transmitted
from the server.

module id message id

Message

Detail