KERN & SOHN CCS Version 2.0 User Manual
Page 38

38
CFS/CCS-BA-e-1220
Add up:
Place weighing goods A.
Wait until the stability display appears, then press
. The weight value or the
number of items are stored and printed out.
Remove the weighed good. More weighed goods can only be added when the
display ≤ zero.
Place goods to be weighed B.
Wait until the stability display appears, then press
. The weight value or the
number of items are added in the summation memory and printed out. Total
weight, number of weighing procedures as well as total parts counting appear 2
sec.
Add more weighed goods as described before. Please note that the balance
must be unloaded between the individual weighing procedures.
This process can be repeated 99 times or until the weighing range of the balance
is exhausted.
Display of the saved weighing data:
Press
, the total weight, number of weighing procedures as well as total
number of items are displayed.
Total weight placed on
balance:
Number of weighing
procedures:
Total number of pieces:
Delete weighing data:
Press
, the total weight, number of weighing procedures as well as total
number of items are displayed. During this display press
. The data in the
summation memory are deleted.