Table linearization points – KERN & SOHN CCS Version 2.0 User Manual
Page 31

CFS/CCS-BA-e-1220
31
If necessary, at balance zero display using
select the weighing unit [kg or lb], which shall be
used for linearisation. The appeared [] displays
the current weighing unit.
Acknowledge with
.
Ensure that there are no objects on the weighing
pan. After successful standstill control press the
button.
The weighing value of the first weight will be
displayed.
Place weight 1 carefully in the centre of the
weighing plate, wait for the stability display and
confirm by
.
(example)
The weighing value of the second weight will be
displayed.
Remove weight 1
Place weight 2 carefully in the centre of the
weighing plate, wait for the stability display and
confirm by
.
(example)
After the adjustment the balance will carry out a
self-test. Remove adjusting weight during selftest,
balance will return into weighing mode
automatically.
In case of fault or a wrong weight an error message
(
) will appear, then repeat the
linearization procedure.
8.1.1 Table Linearization points
Max
Display
Weight 1
Weight
Display
Weight 2
Weight
6 kg
„A 2 kg“
2 kg
„A 6 kg“
6 kg
15 kg
„A 5 kg“
5 kg
„A 15 kg“
15 kg
30 kg
„A 10 kg“
10 kg
„A 30 kg“
30 kg