Linearisation cfs 50k-3 models – KERN & SOHN CCS Version 2.0 User Manual
Page 32

32
CFS/CCS-BA-e-1220
8.2 Linearisation CFS 50K-3 models
Operation
Display
Switch-on balance and during the selftest press
.
With the numerical keys enter the password „
9999“:
Confirm input with
.
„tECH LoCAL“ is displayed, should it not be
displayed, with
select „tECH LoCAL“.
Acknowledge with
.
The appeared [] displays the current balance.
If necessary, at balance zero display using
select the weighing unit [kg or lb], which shall be
used for adjustment. The appeared [] displays the
current weighing unit.
Acknowledge with
.
Ensure that there are no objects on the weighing
pan. After successful standstill control press the
button.
„LoAd 1“ will be displayed
15 kg adjustment weight placed carefully in the
centre of the weighing plate, wait for stability display
and confirm with
.
„LoAd 2“ will be displayed,
30 kg adjustment weight placed carefully in the
centre of the weighing plate, wait for stability display
and confirm with
.