7.3 Adjustment bulk material scales CFS 3K-5, CFS 6K0.1, CFS 15K0.2, CFS
30K0.5
Not pre-configured by KERN
Operation
Display
Switch-on balance and during the selftest press
.
Use the number keys to enter password
:
standard password „0000“
Confirm input by
.
Use
to select bulk material scales („remote“).
The appeared [] displays the current balance.
Acknowledge with
.
If necessary, at balance zero display using
select the weighing unit [kg or lb], which shall be used
for adjustment. The appeared [] displays the current
weighing unit.
Acknowledge with
.
Ensure that there are no objects on the weighing pan.
After successful standstill control press the
button.
Enter value of the necessary adjustment weight using
numeric keyboard and acknowledge with
.
In order to achieve high-quality weighing results in the
sense of the measuring technology, it is recommended
to select the nominal value as high as possible.
We recommend 80 % max.
Place the required adjusting weight carefully in the
centre of the weighing plate and press
to confirm.
After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will
return into weighing mode automatically.
In case of an adjustment error or incorrect adjusting
weight the display will show an error message
(
), repeat adjustment process.