12 error messages, 13 service, maintenance, disposal – KERN & SOHN ALJ 310-4A User Manual
Page 76
ALJ/ALS/PLJ/PLS-BA-e-1338
76
12 Error messages
ERR01
Weight value instable or zeroing not possible. Check the
environmental conditions.
ERR02
Adjustment error, e.g. environmental conditions instable
ERR03
Adjustment error e.g. incorrect adjustment weight}
ERR04
Piece weight too small/instable
ERR05
Data transfer not possible, as weighing value is instable. Check
the environmental conditions.
ERR06
Weighing value in density determining mode instable.
Check the environmental conditions.
ERR07
Incorrect data reading
(piece counting, density determination,…)
ERR08
Error at internal adjustment
“UNLOAD”:
Weighing range not achieved.
Check position of the weighing plate
“CAL But”
Adjust balance
Weighing range exceeded, placed load exceeds the capacity of
the balance. Unload the balance.
Weighing range not reached, e.g. weighing plate not in place
13 Service, maintenance, disposal
13.1 Cleaning
Before cleaning, please disconnect the appliance from the operating voltage.
Please do not use aggressive cleaning agents (solvents or similar agents), but a
cloth dampened with mild soap suds. Ensure that no liquid penetrates into the device
and wipe with a dry soft cloth.
Loose residue sample/powder can be removed carefully with a brush or manual
vacuum cleaner.
Spilled weighing goods must be removed immediately.
13.2 Service, maintenance
The appliance may only be opened by trained service technicians who are authorized
by KERN.
Before opening, disconnect from power supply.
13.3 Disposal
Disposal of packaging and appliance must be carried out by operator according to
valid national or regional law of the location where the appliance is used.