7.1 Models with external weight (KERN ALS/PLS)

The adjustment is locked for verified balances.

7.1.1 Adjustment with recommended adjustment weight (factory setting)

Weight value of the required adjustment weight see chpt. 1 “Technical
specifications“:

 Ensure that there are no objects on the weighing plate. Press the CAL key

 Wait until the weighed value for the required adjustment weight appears

flashing.

During the flashing display put the required adjustment weight carefully in the

center of the weighing plate.
The flashing display disappears.
After successful adjustment the balance automatically returns to weighing mode.

 Take away adjustment weight