10.7 Peak value function

This function displays the highest load value (peak value) of a weighing.
The peak value remains in the display until it will be deleted.

 In weighing mode press MENU

button. The first menu item „count“ is displayed.

 Press MENU button repeatedly

 Confirm using PRINT

button, the taring process is automatically started. „0-t“ is

displayed.
From here the balance is in peak value mode, where

a „M” pops up.

 Load weighing plate. The maximum load value is displayed.

 The peak value remains in the display until the TARE button is pressed. Then

the balance is ready for further measurements.

Return to weighing mode

 Press the ON/OFF key