beautypg.com

KERN & SOHN ALJ 310-4A User Manual

Page 66

background image

ALJ/ALS/PLJ/PLS-BA-e-1338

66

10.5 Percent determination

Percent determination allows weight display in percent, in relation to a reference
weight.

10.5.1 Entering the reference weight by weighing

 In weighing mode press MENU

button. The first menu item „count“ is displayed.

 Press MENU button repeatedly

 Acknowledge using PRINT button, the current setting is displayed.
 Using MENU button select

„PErc A

 If using a weighing container, put this before pressing the PRINT button. The

taring process is automatically started. „0-t“ is displayed.

 Put on reference weight (=100 %)
 Wait until the weight display of the balance has become stable and then press

the PRINT button. The weight is adopted as reference (100%).

 Remove reference weight. From now the balance is in percent determining

mode.

 Place goods to be weighed on balance.

The weight of the sample is displayed in percentage in terms of the reference
weight.