10.5.2 Numeric entering of the reference weight
In weighing mode press MENU
button. The first menu item „count“ is displayed.
Press MENU button repeatedly
Acknowledge using PRINT button, the current setting is displayed.
Using MENU button select
„PErc n“
Confirm using the PRINT button, the display to enter the reference weight
appears.
Press the arrow buttons to enter the reference weight (100%).
Confirm input by pressing the PRINT button. From now the balance is in percent
determining mode.
Place goods to be weighed on balance.
The weight of the sample is displayed in percentage in terms of the reference
weight.
Return to weighing mode