10.1 Parts counting
Before the balance can count parts, it must know the average part weight (i.e.
reference). Proceed by putting on a certain number of the parts to be counted. The
balance determines the total weight and divides it by the number of parts, the so-
called reference quantity. Counting is then carried out on the basis of the calculated
average piece weight.
As a rule:
The higher the reference quantity the higher the counting exactness.
In weighing mode press MENU button. The fi
rst menu item „count“ is displayed.
Acknowledge using PRINT button, the current reference quantity is displayed.
Using the arrow buttons select reference quantity (10, 25, 50, 100 or
manual = numerical input of the reference weight, see chap. 10.1.3).
Scroll forward using MENU button
Scroll backward using CAL button
If using a weighing container, put this before pressing the PRINT button. The
taring process is automatically started. „0-t“ is displayed.
Confirm selected reference number of items by pressing the PRINT button.
Place as many pieces to add-up as required by the set reference piece number.
Acknowledge using PRINT button and wait for standstill control.
After successful reference determination the current quantity is displayed.
Remove reference weight. The balance is now in parts counting mode counting
all units on the weighing plate.