beautypg.com

KERN & SOHN ALJ 310-4A User Manual

Page 59

background image

ALJ/ALS/PLJ/PLS-BA-e-1338

59

The display for weight determination of the "sample in air“ appears.

 Confirm by pressing the PRINT button.
 If necessary, tare and put the sample.
 Wait until the weight display of the weighing balance has become stable and

then press the PRINT button.

„WEI AIR“ appears flashing.

The display for weight determination of the "sample in the measuring fluid“
appears.

 Acknowledge using PRINT button
 Immerse sample holder and tare by pressing the TARE-key
 Place sample and immerse it

Make sure that the sample holder does not touch the beaker.

 Wait until the weight display of the weighing balance has become stable and

then press the PRINT button.

„WEI LIQ“ appears flashing.

 The density of the sample is shown

Return to weighing mode

 Press the ON/OFF key

 or use the MENU button to start a new measuring cycle.

If at the density determination errors have appeared, „d-----„ is displayed.

This manual is related to the following products: