KERN & SOHN ALJ 310-4A User Manual
Page 6
ALJ/ALS/PLJ/PLS-BA-e-1338
6
KERN
ALS 160-4A
ALS 250-4A
Weighing range (max)
160 g
250 g
Readability (d)
0.1 mg
0.1 mg
Reproducibility
0.1 mg
0.1 mg
Linearity
± 0.3 mg
± 0.3 mg
Stabilization time (typical)
4 sec.
4 sec.
Minimum unit weight at
piece counting
> 0.5 mg
> 0.5 mg
Warm-up time
8 hours
Recommended adjustment weight,
not added (class)
150 g (E2)
200 g (E2)
Reference quantities at
piece counting
10, 25, 50, 100 freely selectable
Weighing Units
ct, g, gn, lb, mo, oz, ozt,
tl (HK), tl (Singap. Malays), tl (Tw), pen
Electric Supply
24 V AC, 500 mA
Operating temperature
+ 5° C …. + 35° C
Humidity of air
max. 80 % (not condensing)
Housing (B x D x H) mm
210 x 340 x 330
Dimensions of windshield
(B x D x H) mm
160 x 140 x 205 (inside)
190 x 195 x 225 (outside)
Weighing plate (stainless steel)
Ø 80 mm
Weight kg (net)
5.7 kg
Interface
RS 232C