Manual MSC-64: Modbus on Ethernet
5.10 Save calibration, setup and setpoint parameters Commands – CS, WP, SS
The setup and calibration parameters can be divided into 3 groups:
• Calibration parameters: CZ, CG, DS, DP & ZT are saved by the CS command. 
• Setup parameters (other than setpoint): FL, FM, NR, NT, BR, AD, etc. are saved by the WP 
command.
• Setpoint parameters: Sn, Hn and An are saved by the SS command.
 
 
CS
Save the calibration parameters
[ Register = 2066, Type = Int16, Value = 004 ]
Note: Calibration parameters can only be saved if the TAC code is known and precedes the CS 
command. See the CE and CS commands on 
page 26.
Both the setup parameters and the setpoint parameters are stored in EEPROM using the WP 
and SS commands respectively. 
 
 
WP
Save the setup parameters
[ Register = 2066, Type = Int16, Value = 008 ]
With this command the settings of the “Filter” (FL, FM) , the “No-Motion” (NR, NT) and the 
communication (AD, BR) will be saved in the EEPROM.
Master (PC / PLC) sends
Device responds
Result
WP OK
Parameter saved
WP ERR
Error
 
 
SS
Save the setpoint set-up parameters
[ Register = 2066, Type = Int16, Value = 032 ]
With this command the settings of the setpoints (Sn), the “setpoint hysteresis” (Hn) and the 
“setpoint action” (An) will be saved in the EEPROM.
Master (PC / PLC) sends
Device responds
Result
SS OK
Parameter saved
SS ERR Error