Flintec MCS-64 ModbusOnEthernet User Manual
Page 29
Manual MCS-64: Modbus on Ethernet
Page 29
TL Trigger
Level
[ Register = 2400, Type = Int32 ]
[ Register = 211E, Type = Int32 ]
Set the trigger level for rising edge start of measurement. Permitted values are in the range
0…262143.
Master (PC / PLC) sends
Device responds
Result
TL T+99999
Request: TL=99999
TL_1000 OK
Setting: TL=1000
With regard to the trigger commands SD and MT, a check weighing will automatically start when
the weight overshoots by e.g. 1.000d (increments), e.g. 100,0 g.
Factory default: 199999
TE
Trigger Edge
[ Register = 2402, Type = Int32 ]
[ Register = 211C, Type = Int32 ]
Issuing the TE command selects rising or falling edge trigger. Parameter = 0 selects falling edge
and parameter = 1 selects rising edge.
Master (PC / PLC) sends
Device responds
Result
TE E:001
Request: TE=1
TE_0 OK
Setting: TE=0
Factory default: 0 [falling edge]
Note: This command cannot be used together with the TL command.
TR
Trigger
[ Register = 2062, Type = Int16, Value = LDM# ]
Cancel measurement
[ Register = 2062, Type = Int16, Value = LDM# + 128 ]
[This command will start the measuring cycle on any LDM# (independent from the LDM selector
register) in the same way as the hardware trigger input.]
Master (PC / PLC) sends
Device responds
Result
TR OK
Trigger started
Note: This function can be used as a soft trigger in a check weigher application (firmware
88.183) or to start a filling process (firmware 88.184)