Flintec MCS-64 ModbusOnEthernet User Manual
Page 32
Page
32
Manual MSC-64: Modbus on Ethernet
TW
Tare Window
[ Register = 240A, Type = Int32 ]
Set the Tare Window in counts (digits) without decimal point. Tare window (TW) allows an
automatic Tare update. If TW = 0 this function is not active. If TW = 100, this means a new tare
value will be taken when the net average weight of an empty scale is within 100 counts or
division of zero. The new average tare value is calculated over the average tare time defined by
TI. If the tare average is outside tare window, the tare will not be updated.
Master (PC / PLC) sends
Device responds
Result
TW T+00000
Request: TW=00000
TW_100 OK
Setting: TW=100d
Factory default: 00000
TI
Tare Time
[ Register = 240C, Type = Int32 ]
Set the Tare Time in milliseconds [ms].
During the Tare Time a “tare-average” will be calculated
by the system.
Master (PC / PLC) sends
Device responds
Result
TI T+00000
Request: TI=00000
TI_200 OK
Setting: TI=200ms
Factory default: 00000
HT
Hold Time
[ Register = 2408, Type = Int32 ]
Set the Hold Time in milliseconds [ms].
During the Hold Time the weight value must be over
setpoint that a digital output can be switched. This means that a short signal peak will not lead to
the switching of a digital output.
Master (PC / PLC) sends
Device responds
Result
HT T+00000
Request: HT=00000
HT_200 OK
Setting: HT=200ms
Factory default: 00000