Manual MCS-64: Modbus on Ethernet

Page 17

DS

Display step size

[ Register = 2216, Type = Int32 ]

This command allows the output to step up or down by a step size other than 1. Permitted values
are 1, 2, 5, 10, 20, 50, 100 and 200.

Master (PC / PLC) sends

Device responds

Result

DS S+00002

Request: display step size 2

CE

E+00017 (example)

Request: TAC-counter CE 17

CE_17 OK

Calibration sequence active

DS_50 OK

Setting: DS = 50


DP

Set decimal point position

[ Register = 2214, Type = Int32 ]

This command allows the decimal point to be positioned anywhere between leftmost and right-
most digits of the 5-digit output result. Position 0 means no decimal point.

Master (PC / PLC) sends

Device responds

Result

DP P+00002

Request : position of dec. point

CE

E+00017 (example)

Request: TAC-counter CE 17

CE_17 OK

Calibration sequence active

DP_0 OK

Setting: no decimal point


CZ

Set calibration zero point

[ Register = 2212, Type = Int32 ]

This is the reference point for all weight calculations, and is subject to TAC control.

Master (PC / PLC) sends

Device responds

Result

CE

E+00017 (example)

Request: TAC-counter CE 17

CE_17 OK

Calibration sequence active

CZ OK

Zero point set

Factory default: approx. 0 mV/V input signal


CG

Set calibration gain (span) value

[ Register = 2206, Type = Int32 ]

This is the reference point for calibration under load, and is subject to TAC control.
Permitted values are 1…262143.

Master (PC / PLC) sends

Device responds

Result

CG G+20000

Request : span 20000d

CE

E+00017 (example)

Request: TAC-counter CE 17

CE_17 OK

Calibration sequence active

CG_50000 OK

Setting: span 50000d

Factory default: 200000 d = 2.000 mV/V input signal = 20 kg

For calibration an input signal near the display maximum (CM) will give the best system
performance. The minimum calibration load of at least 20% is recommended.