Manual MCS-64: Modbus on Ethernet

Page 21

5.4 Filter setting Commands – FM, FL, UR

Using the commands FM and FL, a digital filter type and strength can be set which will eliminate
most of the unwanted disturbances. The command UR is used for the average building. Please
note that these filters are positioned immediately after the A/D Converter and therefore affect all
aspects of the weighing operation.


FM

Filter Mode FIR / IIR

[Register = 2110, Type = Int32 ]

Choose filter mode, permitted values are “0” for IIR and “1” for FIR.

Master (PC / PLC) sends

Device responds

Result

FM M+00001

Request: FM = 1 (FIR)

FM_0 OK

Setting: FM = 0 (IIR)

WP OK

Setting saved

The digital IIR filter works as a low-pass filter of 2

nd

order with Gaussian characteristic, damping

is 40 dB/decade; see table mode 0. The digital FIR filter works as a low-pass filter with quick
response; damping see table mode 1. Factory default: 0


FL

Filter settings

[Register = 2106, Type = Int32 ]

Command for setup cut off frequency, permitted values are 0 … 8.

Master (PC / PLC) sends

Device responds

Result

FL F+00003

Request: FL = 3

FL_1 OK

Setting: FL = 1

WP OK

Setting saved

Filter values can be chosen between 0 and 8, see table below.
FL= 0 means no filter in mode 0 or 1 (command FM).
Factory default: 3

Mode 0 Characteristic (IIR-Filter)

FL

Settling time to

0.1%

(ms)

3dB Cut-off

frequency

(Hz)

Damping

@300Hz

(dB)

Output-rate*

(samples/s)

0 no

filtering

**

600

1 55

18

57

600

2 122

8

78

600

3 242

4

96

600

4 322

3

104

600

5 482

2

114

600

6 963

1

132

600

7 1923 0.5 149 600
8 3847 0.25 164 600

* Output-rate = 600/2

UR

samples/s

** Antialiasing filter 17 Hz @ 60 dB/dec