FLUKE 900 SERVICE MANUAL
Maintenance
TABLE 1
Offset Shift
Fluke P/N
R
4 0
80
120 160 200 240
365k
0
0
0
1
1
1
921168
150k
1
1
2
2
3
3
921171
113k
1
2
2
3
4
4
921176
8 6k
2
2
3
4
5
5
921184
57k
3
4
5
6
6
7
921189
47k
4
5
6
6
7
8
921192
TABLE 2
Offset Shift
Fluke P/N
C
40
80
120 160 200 240
1. OpF
1
1
1
2
2
2
921197
1.5pF
1
1
2
2
3
3
921200
1.8pF
1
1
2
3
3
3
921205
2.2pF
2
2
3
3
4
4
921213
3.3pF
2
3
3
4
4
5
921218
4.7pF
2
3
4
5
6
6
921221
6.8pF
3
4
6
6
7
8
921226
NOTE: All the offsets in Table 2 are meant as negative (i.e. they must be subtracted from
the values in the calibration data printout).